This is a fully functional Hirayama HAST PC-422R8 Highly Accelerated Stress Test System chamber in excellent cosmetic and working condition. This PC-422R8 is designed to perform both saturated and unsaturated accelerated humidity/temperature/pressure stress testing under a temperature range 105.0 -150.0 °C, humidity range of 65%RH – 100%RH, and pressure range of 0.019 – 0.208 MPa.
This HAST chamber is easy to use with color touch panel /screen integrates various button operations and displays for monitor, graph display, mode setting, program setting, test name. Different temperature, humidity, and pressure conditions have been tested, including commonly used settings such as 121°C / 100% RH (Autoclave), 130°C / 85% RH (HAST/BIAS, UHAST). Please see the photos and video for details.
This chamber has two control methods installed, HUM / Unsaturation control and STD / Saturation control. The unsaturation control mode prevents dew condensation and can set the humidity at any value between 65%RH and 100%RH that meets the test conditions defined in IEC 60068-2-66 and JIS C 60068-2-66. Saturation control, also called wet saturation test, is a 100%RH saturation test mode with humidity maintained at 100%RH from a start.
The PC-422R8 system has three exhaust modes (slow cooling with humidity maintained mode, cooling with moisture kept mode, and rapid exhaust mode) and two operation modes (fixed-value operation and programmed operation).
20 voltage application terminals are arranged on the inner face of the door. Connection work with specimens at a voltage application test is executed outside of the chamber. The RS232C port is used as the communication interface for centralized monitoring and centralized control. In addition, an electric clamp system with high safety performance is used for door locking and release.
Typical Applications
• Semiconductor package reliability (ICs, BGA, QFN, CSP, etc.)
• Automotive electronics qualification
• PCB and electronic component moisture resistance testing
• Accelerated life and failure analysis testing
Key Features
• Saturated and unsaturated humidity operation
• High-accuracy temperature and humidity control
• Dual-vessel construction for stable environmental performance
• Stainless steel pressure chamber for corrosion resistance
• Programmable test profiles and timer operation• Safety interlocks and pressure protection
• Bias test support with voltage application terminals
This chamber runs on 220V single-phase, 60Hz, 15A power. It is a complete, ready-to-use system. With an order in place, it can be shipped out the next business day.
Technical support and training will be available after the chamber is delivered.
Please see all the photos and videos for details.
Specifications:
Condition: Used
Product: Hirayama / HAST PC-422R8 Highly Accelerated Stress Test System Chamber
Model: PC-422R8Working
Space: 3 cu. ft. / 84L (φ16.5” x D25.9”)
Effective chamber size: 1.4 cu. ft. /40L (φ13.4” x D18.7”)
Temperature
• Range:
• 105.0 – 133.3 °C @ 100 %RH
• 110.0 – 140.0 °C @ 85 %RH
• 118.0 – 150.0 °C @ 65 %RH
• Control Method: PID control with SSD drive
• Control Accuracy: ±0.5 °C
• Distribution Accuracy: ±0.5 °C @ 100 %RH ±1.0 °C @ 85 %RH
Humidity
• Range: 65 %RH – 100 %RH
• Control Accuracy: ±3 %RH (@ 85 %RH)
Pressure
• Range: 0.019 – 0.208 MPa
Programmable Controller: Color touchscreen - Fixed value mode and Program mode (multi-step profiles)
Temperature Rise Time: ~ 70 min (time to 120℃ 85%RH from room temperature)
Exhaust Mode: EXHT MODE selector switch
MODE1: Slow cooling with humidity maintained mode
MODE2: Cooling with moisture kept mode
MODE3: Rapid exhaust mode
Bias Testing /Voltage Application Terminal: 20 (Applied voltage: 125V, within 30W in total)
Water Supply Method: Built-in water supply tank automatically feeding the steam generator
Door Open-close Method: Drawer door structure, Electric clamp method
Communication Interface: RS-232C for PC monitoring and data logging
Safety Devices and Alarm Functions: Pressure safety valve, Over-pressure prevention device, Overheat prevention device, Low water cut off device, Water supply abnormality detection device, Over-current and earth leakage breaker, Circuit breaker for heater, Fuse, Thermal fuse, Door safety system (Door lock check mechanism during operation, Door open / close check mechanism), Protection for settings, Hot water drainage treatment system, Self-diagnosis function of sequence controller, Negative pressure prevention system
Compliance & Test Standards:
JEDEC JESD22-A110 – Highly Accelerated Temperature and Humidity Stress Test (HAST)
JEDEC JESD22-A118 – Accelerated Moisture Resistance, Unbiased HAST (Air-HAST)
JEDEC JESD22-A101 / A102 (related humidity stress methods)
Power: 220 V, 1Ph, 60 Hz, 15A
Warranty : 60-Day. See full term.